Cleaning and steam sterilization
General guidance for stainless hand instruments. Always follow your local infection-control rules and your autoclave manufacturer's instructions.
Before first use
Instruments are supplied non-sterile. Clean, inspect and sterilize every new instrument before it is used on a patient.
Reprocessing sequence
- Point of use. Wipe off gross debris and keep instruments moist so blood does not dry on.
- Cleaning. Use an ultrasonic bath or washer-disinfector with a neutral or mildly alkaline instrument detergent. Open hinged instruments.
- Rinse. Use softened, deionised or distilled water to avoid spotting.
- Dry and inspect. Check tips, edges and joints under good light. Lubricate hinges with a steam-permeable instrument lubricant.
- Package. Use pouches or cassettes, with hinged instruments unlocked.
- Sterilize. Steam sterilize. Commonly used cycles are 134 °C for at least 3 minutes or 121 °C for at least 15 minutes; follow the validated cycle for your autoclave.
- Dry and store. Allow the full drying phase, then store in a clean, dry place.
Staining and spotting
| You see | Likely cause |
|---|---|
| White or rainbow spots | Minerals in rinse or steam water |
| Orange-brown rust spots | Residue left before sterilizing, wet storage, or rust transferred from another instrument |
| Pitting | Chlorides, usually dried blood or saline |
| Black or dark staining | Detergent residue or unsuitable chemicals |
Instruments that are pitted, cracked or misaligned should be taken out of service. Edged working ends can be sent back for free resharpening.